Power Supply | Rated voltage | 100-240 V AC; within ±10% | |
---|---|---|---|
Rated frequency | 50/60 Hz; within ±1 Hz | ||
Rated input | 400 VA | ||
Size | Dimensions (maximum) | 383 (W) × 199 (H) × 506 (D) mm | |
Weight | 19.3 kg | ||
Observation | Analog signal output | VBS composite and Y/C; simultaneous outputs possible. | |
Digital signal output | HD-SDI (SMPTE292M), DVI (WUXGA, 1080 pixels, or SXGA can be selected) | ||
Electronic zoom | The image enlargement level can be selected. 3 modes (1.0×, 1.2×, 1.5×) | ||
Optical-digital observation | The optical-digital observation can be performed. The endoscope compatible with the optical-digital observation is required. | ||
NBI observation | This observation mode uses narrow-band light | ||
IR observation | This observation mode uses infrared light | ||
Documentation | Remote control | The following ancillary equipment can be controlled (specified models only): • Portable memory / • Video recorder / • Video printer / • Image filing system |
|
Recording format and number of recorded images in internal memory | TIFF: no compression | Approx. 120 images | |
JPEG (1/5): Approx. 1/5 compression |
Approx. 636 images | ||
JPEG (1/10): Approx. 1/10 compression |
Approx. 1108 images | ||
These are the numbers of recorded images when both HDTV and SDTV images are recorded. These numbers vary depending on the images. |
|||
Illumination | Examination lamp | LED | |
Cooling | Forced-air cooling | ||
Providing maximum light intensity | Less than 5.64 W (approximately same as Olympus light source that is mounted xenon short-arc lamp 300 W) | ||
Observation mode | WLI or NBI observation | ||
IR observation (when connected to CLV-S200-IR) | |||
Automatic Brightness Adjustment | Automatic brightness adjustment method | LED drive current control | |
Automatic exposure | 17 steps | ||
Brightness mode | Auto | ||
Manual | |||
Classification (Medical Electrical Equipment) | Type of protection against electric shock | Class I | |
Degree of protection against electric shock of applied part | Depends on applied part. Also refer to applied part (camera head or videoscope). | ||
Degree of protection against explosion | The video system center should be kept away from flammable gases |